| E151 Internal PFC Error | ||||
Prepared by: EH | Approved by:
| Valid from: 280426 | Document no.:
| Rev. no.
| |
1. Error Overview
Definition: This error indicates that the generator is reporting an internal power issue.
2. Symptom Checklist
- E151 is displayed on the iPM generator and external controllers.
3. Potential Root Causes
- Occurs when the input voltage falls below 200 Volts, preventing proper power factor correction.
- There is a missing phase to the power connection of the genertor.
If the input voltage is stable and the error persists, an internal failure on the iPM Power Board may be the cause.
- In rare cases, some iPM 100 and iPM 200 generators may falsely report E17 error due to a software limitation.
4. Step-by-Step Resolution
- Check main power of the generator is at the correct voltage.
- Check that each phase is present to the main power.
- Update PU firmware in the generator
5. Escalation
- If the above resolutions did not solve you issue please reach out to our support team. Replacement parts will be needed to resolve this issue.